Vaisala HMP240 User Manual
Page 128

HMP240
Appendix 7: Re-gaining
M210300en
126
1.2.3 CRI
Changing re-gaining parameters
CRI
With command CRI the values of the re-gaining parameters can be changed.
Example: changing the re-gaining interval to 480 min and the settling time
after sensor heating to 360 s
>cri
Interval min : 720 ? 480
Duration s : 120 ?
Temp diff : 0.50 ?
Settl time s : 240 ? 360
k % : 95.00 ?
>
NOTE
To activate new parameter settings immediately,
command RESET must be given. Otherwise the
transmitter starts the re-gaining with new parameters
only when the old re-gaining interval is completed.
1.2.4 FST
Displaying re-gaining status
FST nnn
nnn = ON or OFF
With FST ON the status of the re-gaining is output at the beginning of the
output line.
See also other documents in the category Vaisala Humidifiers:
- Calibration of Digital Transmitters with HMI41 (36 pages)
- Calibration of Series HMDW2030 and HMP130 Transmitter with HMI41 (14 pages)
- Calibration of Series HMDW6070 and HMP140 Transmitter with HMI41 (30 pages)
- HM34 (30 pages)
- HM40 (47 pages)
- HM44 (52 pages)
- HM70 (83 pages)
- HMD40 (1 page)
- HMD60 (4 pages)
- HMD70 (18 pages)
- HMDW110 (62 pages)
- HMDW80 (51 pages)
- HMI41 (74 pages)
- HMP41 (72 pages)
- HMK15 (39 pages)
- HMM100 (71 pages)
- HMM105 (23 pages)
- HMM211 (42 pages)
- HMM212 (36 pages)
- HMM213 (52 pages)
- HMP140 (28 pages)
- HMP155 (84 pages)
- HMP228 (115 pages)
- HMP230 (163 pages)
- HMP260 (118 pages)
- HMP60 (71 pages)
- HMT100 (52 pages)
- HMT120 (87 pages)
- HMT130 (95 pages)
- HMT140 (76 pages)
- HMT310 (88 pages)
- HMT310 (105 pages)
- HMT330 (209 pages)
- HMT360 (97 pages)
- HMT360 (63 pages)
- HMT360N (110 pages)
- HMW40 (1 page)
- HMW90 (110 pages)
- SHM40 (68 pages)
- RDP100 (14 pages)