Preparing the KeepSafe Deluxe for Use

Battery Installation:

The battery-operated KeepSafe Deluxe is portable and long lasting. Fresh alkaline batteries have an

estimated life of 30 days of daily use. Actual life depends on alarm mode, tone and volume you select.
Alarm will “chirp” about every 15 seconds when new batteries are needed. Change batteries at once.
1. If your alarm does NoT have an oN/oFF switch (Cat. 8374NP),

connect a sensor before inserting the batteries. If your alarm DoES

have an oN/oFF switch (Cat. 8374), slide the power switch to the oFF

position before inserting/changing batteries.

2. Press down on the arrow and slide the battery compartment door

completely off (fig. 1). Set battery door aside.

3. Drape the red ribbon across the battery compartment and insert four

(4) new “AAA” batteries as pictured inside the battery compartment

(fig. 2). Take care not to damage battery contacts. Batteries should be

inserted on top of red ribbon and the loose end of the ribbon should

stick out the other end for easy battery removal (fig. 3).

4. lay the loose end of the red ribbon on top of the batteries and reattach

the battery compartment door. Slide it shut, locking it into place.

• Take care when installing new batteries. The alarm will not work if

batteries are installed improperly.

• ALWAYS install a completely new set of batteries when the

chirping due to low battery power sound begins. DO NOT replace

a single cell, but all cells in the alarm. DO NOT mix old and new

batteries, or battery brands within a battery pack (4 batteries).

use of mixed batteries or batteries installed incorrectly may cause

battery damage, and may damage the alarm. Remove any alarm

from use and send to the appropriate facility authority if batteries

are damaged or corroded or the battery compartment has signs of

previous battery corrosion such as white powder residue.

• Batteries can explode or leak and cause damage to alarm if

installed incorrectly, fully discharged, or exposed to liquid, fire

or high temperatures. if battery damage has occurred, or you see any

corrosion, remove the alarm from use immEDiaTEly. DO NOT use the

alarm if battery damage has been detected.

• After changing batteries, test the alarm and sensor for proper operation prior to putting in service

with a patient, and each time before leaving the patient unattended. if the alarm and/or sensor do

not function properly, remove the alarm and sensor from service and replace them with a properly

functioning alarm and/or sensor (see pages 19-20).