Adjusting alarm volume, Keepsafe, Deluxe 11 – Posey KeepSafe® Deluxe User Manual
Page 11
KeepSafe
®
Deluxe
11
Adjusting Alarm Volume
There are two (2) alarm volume settings: low and high. These allow you to select the right volume for
facility and/or patient needs.
All volume settings are within oSHA standards. For maximum staff alert in noisy areas, use HIGHEST
(loudest) volume setting.
alWayS check to ensure staff can hear alarm at the furthest possible distance before
leaving patient unattended.
NEVER place alarm closer than two feet from patient’s ear. Doing so may cause hearing
loss or other injury. For more information, see: OSha OCCuPaTiONal NOiSE EXPOSuRE STaNDaRDS
1910.95.
To Change Volume:
1. Ensure alarm is on and either sensor pad, chair belt sensor or PIR
sensor is attached to alarm. Ensure pressure is on sensor pad, chair belt
sensor is connected, or PIR sensor is connected.
2. Remove the battery compartment door by sliding down to gain access
to the programming buttons.
3. Press VolUME button (fig. 6) to scroll through both the volume
selections.
4. Alarm mode must be set to either “Voice and Tone” or “Tone” (see
page 9) in order to hear the selection.
5. Each time you press the button, the volume changes and a two (2)
second sample plays.
6. Continue to press the VolUME button until you hear the desired volume. The last sample heard is the
volume in use.
7. Replace the battery door and slide shut, locking into place.
NOTE: Cat. 8374NP does not have an oN/oFF switch. To set Alarm Volume for Cat. 8374NP, follow all
steps above except step 1.
Fig. 6
Tone
Volume
Voice
&
Tone
Voice
Tone
Qt
y
4A
AA
A
lk
ali
ne
AN
SI-
24
A I
EC
-L
R0
3