SENSe:FILTer:STATe
Description:
Set or return the current setting of the integration filter on the selected channel. OFF provides no
filtering, and can be used at high signal levels when absolute minimum settling time is required.
ON allows a user-specified integration time, from 50 milliseconds to 20 seconds (see
SENSe:FILTer:TIMe
command). Note that setting the filter time will force the state to ON.
t of filtering, which is set automatically by the power meter based on
that gives a good compromise between measurement noise and
SENSe[1|2]:FILTer:STATe
Argument:
SENSe:FILTer:TIMe
Description:
rn the current length of the integration filter on the selected channel. If the filter state is
d if set to OFF, a time query will return 0.00.
ON.
meric value>
econds in 50 millisecond increments
AUTO uses a variable amoun
the current signal level to a value
settling time at most levels.
Syntax:
Set or retu
set to AUTO, querying the time will return -0.01, an
Note that setting the filter time will force the state to
Syntax: SENSe[1|2]:FILTer:TIMe Argument: Remote Operation 5-25