10 read queries, Read:cw:power, 10 read queries -23 – Boonton 4240 RF Power Meter User Manual
Page 83: Read:cw:power? -23, Boonton 4240 series rf power meter, Remote operation 5-23

Boonton 4240 Series RF Power Meter
5.5.10 READ Queries
he purp
AD
ata. READ
s the
is equivalent to ABORtin
ready. READ? generally
every time it is issued, RE
this case, use FETCh que
SENSe:FILTer
) is filled.
r?
Description:
Return current average amplitude reading in channel units.
READ[1|2]:CW:POWer?
T
ose of the RE
? group of queries is to initiate a measurement cycle, acquire data, and return specific measurement
d
? perform
initiation, data acquisition, postprocessing, and data output portions of the measurement. READ?
g any operation in progress, INITiating a new measurement, then FETChing the data when it is
does not return data unless acquisition is complete. Since READ? INITiates a new measurement
AD? queries should not be used for free running data acquisition (INITiate:CONTinuous ON) - in
ries instead. The measurement is generally considered complete when the integration filter (see
READ:CW:POWe
Syntax:
Returns:
CC, Average power (watts, dBm)
Where CC is the measurement condition code.
Remote Operation
5-23