5.5.11 SENSe Subsystem
The purpose of the SENSe command subsystem is to directly configure device specific settings used to make measurements,
generally parameters related to the RF power sensor and signal processing. The SENSe commands are used to configure the
power meter for acquiring data. SENSe enables you to change measurement parameters such as filtering or averaging,
perating frequency and calfactors, and measurement gain or offset. The numeric suffix of the SENSe program mnemonic in
e SENSe commands refers to a hardware measurement “channel”, that is SENSe1 and SENSe2 represent the instrument’s
paths, respectively. The SENSe commands generally DO NOT affect the data processing
nd display portion of the measurement (see the CALCulate subsystem, Section 5.5.2). Note that SENSe2 commands will
nerate
d w
ENSe:CORRe ion:CALFacto
Descript
:
t or r
lfact
ote setting a
lfactor
ll o
as calculated
d app
erating
operating frequency will
erride
ing.
Syntax:
ction:CALFa
Argume
meri
.00 to 3.00 dB
Description:
r re
in use on the selected channel.
Syntax:
Se[
Rection:DCYC
t:
percent
ENSe:CORRection:FREQuency
or return the RF frequency for the current sensor, and apply the appropriate frequency
Syntax: SENSe[1|2]:CORRection:FREQuency
t:
l sensor may have narrower range)
tion:OFFSet
Description:
Set or return a measurement offset in dB for the selected sensor. This is used to compensate for
external couplers, attenuators or amplifiers in the RF signal path ahead of the power sensor.
Syntax: SENSe[1|2]:CORRection:OFFSet
Argument:
o
th
SENSOR 1 and SENSOR 2 signal
a
ge
an error if use
ith a single channel 4240 Series instrument.
S
ct
r
ion
Se
eturn the frequency ca
or currently in use on the selected channel. N
nd wi
lfactor that w
ca
with this comma
verride the “automatic” frequency ca
an
lied when the op
frequency was set, and setting the
sett
ov
this calfactor
SENSe[1|2]:CORRe
ctor
nt:
c value> = –3 SENSe:CORRection:DCYCle Set o turn the duty cycle correction factor currently SEN 1|2]:COR le Argumen S Description: Set calfactor from the sensor’s EEPROM table. Application of this calfactor cancels out the effect of Argumen SENSe:CORRec Remote Operation 5-24
variations in the flatness of the sensor’s frequency response. If an explicit calfactor has been set,
either manually or via the SENSe:CORRection:CALFactor command, entering a new frequency
will override this calfactor and use only the “automatic” frequency calfactor.