Problem probable cause solution – AirSep MN138-1 H User Manual
Page 36

30
MN138-1 Rev H
02/14
Problem
Probable Cause
Solution
Low oxygen concentration.
•
Ambient or unit’s temperature is too
high.
•
Leak.
•
Reduced air intake (suction)
•
Restriction in exhaust muffler
•
Contaminated sieve beds
•
Weak compressor.
•
Faulty circuit board.
•
Faulty solenoid valve
•
Unit operating above temperature
range specifications.
•
Leak test and repair.
•
Blocked air intake or exhaust.
•
Defective cabinet fan.
•
Check compressor intake path for
obstruction. Clean/remove
obstruction.
•
Replace or clean exhaust muffler.
•
Replace sieve beds.
•
Check system pressure, and rebuild or
replace compressor.
•
Replace circuit board
•
Replace valve manifold.
No flow indicated on flowmeter,
and 6-beep no flow alarm is
activated.
•
Flowmeter is turned off.
•
Regulator is turned off.
•
Cannula tubing blocked or kinked
•
External restriction at outlet of unit
•
Internal restriction of airflow.
•
Turn flowmeter knob
counterclockwise to allow flow of
oxygen.
•
Unlock regulator knob and turn
clockwise to allow flow of oxygen.
•
(Refer to section 4.8 Product
Regulator
•
Check and Setting for proper
adjustment).
•
Check cannula tubing. Replace if
necessary.
•
Reduce/remove attachments from
outlet of unit (humidifier bottle,
tubing, etc).
•
Check internal tubing for blockage or
kinks.
Alarm does not sound.
•
Unit has been turned off for an
extended period of time.
•
Faulty electrical connection.
•
Faulty buzzer.
•
Faulty I/0 power switch.
•
Faulty circuit board.
•
Turn on unit and let run for several
minutes to charge alarm capacitor.
With unit running, disconnect from
power source and verify power failure
alarm (1-beep).
•
Repair electrical connection
•
Replace buzzer.
•
Replace I/0 power switch.
•
Replace circuit board.