beautypg.com

Visionaire™ en, En visionaire – AirSep MN137-1 D User Manual

Page 84

background image

VisionAire™

EN

AirSep

®

Corporation

EN-27

MN137-1 rev. D

Recommended separation distances between portable and mobile RF communications equipment and the VisionAire

The VisionAire is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the VisionAire can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the VisionAire as recommended below, according to the
maximum output power of the communications equipment.

Rated maximum

output power of

transmitter

W

Separation distance according to frequency of transmitter

(m)

from 150kHz to 80MHz

d= 1.2 x

√P

from 80MHz to 800MHz

d= 1.2 x

√P

from 800MHz to 2.5GHz

d= 2.3 x

√P

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.

Guidance and manufacturer’s declaration – electromagnetic emissions

The VisionAire is intended for use in the electromagnetic environment specified below. The customer or the user of the VisionAire
should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic Environment guidance

RF emissions

CISPR 11

Group 1

The VisionAire uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.

RF emissions

CISPR 11

Class B

The VisionAire is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-
voltage power supply network that supplies buildings
used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/flicker emissions

IEC 61000-3-3

Complies

EN

VisionAire™

EN-28

AirSep

®

Corporation

MN137-1 rev. D

7.0

Symbols/Abbreviations


Symbols are frequently used on equipment in preference to words with the
intention of lessening any possibility of misunderstanding caused by language
differences. Symbols can also permit easier comprehension of a concept within a
restricted space.

The following table is a list of symbols and definitions that may be used with
AirSep’s VisionAire Oxygen Concentrator. These symbols are referenced from
the appropriate International Electro-technical Commission (IEC) standards:

Symbol

Description

Symbol

Description

ON (power switch on)

OFF (power switch off)

No smoking

Do not disassemble

Type B equipment

Consult instructions for use

Warning – Describes a

hazard or unsafe practice that

if not avoided can result in

severe bodily injury, death or

property damage

Class II Device

Caution – Describes a hazard

or unsafe practice that if not

avoided can result in minor

bodily injury or property

damage

Complies with the 93/42/EEC

directive drawn up by the

approved organization No.

0459

Note – Provides information

important enough to

emphasize or repeat

Safety agency for CAN/CSA

C22.2 No. 601.1 M90 for

medical electrical equipment

Consult the accompanying

documents

Keep unit and accessories dry

Use no oil or grease

Proper disposal of waste of

electrical and electronic

equipment required