Problem

Probable Cause

Solution

All other problems.

Set I/0 power switch to the
0 position, use your reserve
oxygen supply (if provided), and
consult your Equipment Provider
immediately.

Oxygen
Concentration
Indicator remains lit
(amber)

Oxygen concentration
is

82% (±3%).

Set I/0 power switch to the
0 position, use your reserve
oxygen supply (if provided), and
consult your Equipment Provider
immediately.

EN

VisionAire™

EN-24

AirSep

®

Corporation

MN137-1 rev. D

6.0

Product Specifications

VisionAire

Oxygen Concentrator

Flow Specifications

(1)

1 lpm – 5 lpm

±10% of indicated setting, or 200 mL,
whichever is greater

(1)

Oxygen Concentration

(1)

90% (+ 5.5% / – 3%)

Electrical Requirements

115 VAC / 60 Hz, 3.0 A
230 VAC / 50 Hz, 1.5 A
230 VAC / 60 Hz, 1.5 A

Power Consumption

290 Watts

Sound

40 dBa

Dimensions

14.1 in. W x 11.5 in. D x 20.8 in. H

(35.8 cm W x 29.2 cm D x 52.8 cm H)

Weight

30 lbs (13.6 kg)

Electrical Shock Protection

Class II, Type B

Environmental Limit

Conditions

(2)

Operating conditions: 5

o

C to 40

o

C

(41

o

F to 105

o

F ) at altitudes up to 10,000 ft.

(523mmHg) above sea level.
Storage: -20

o

C – 60

o

C (-4

°F – 140°F)

Relative Humidity: up to 95% RH
(non – condensing)

(1) Based on 21

°C (70°F) at a nominal operating pressure range of

0-7 kPa (0-1 psig back pressure). Maximum output pressure: 55 kPa
(7.98 psig)

(2) Operating outside of these operational specifications can limit the

concentrator's ability to meet Oxygen Concentration specification at
higher liter flow rates.