Problem
Probable Cause
Solution
All other problems.
Set I/0 power switch to the
0 position, use your reserve
oxygen supply (if provided), and
consult your Equipment Provider
immediately.
Oxygen
Concentration
Indicator remains lit
(amber)
Oxygen concentration
is
≤
82% (±3%).
Set I/0 power switch to the
0 position, use your reserve
oxygen supply (if provided), and
consult your Equipment Provider
immediately.
EN
VisionAire™
EN-24
AirSep
®
Corporation
MN137-1 rev. D
6.0
Product Specifications
VisionAire
™
Oxygen Concentrator
Flow Specifications
(1)
1 lpm – 5 lpm
±10% of indicated setting, or 200 mL,
whichever is greater
(1)
Oxygen Concentration
(1)
90% (+ 5.5% / – 3%)
Electrical Requirements
115 VAC / 60 Hz, 3.0 A
230 VAC / 50 Hz, 1.5 A
230 VAC / 60 Hz, 1.5 A
Power Consumption
290 Watts
Sound
40 dBa
Dimensions
14.1 in. W x 11.5 in. D x 20.8 in. H
(35.8 cm W x 29.2 cm D x 52.8 cm H)
Weight
30 lbs (13.6 kg)
Electrical Shock Protection
Class II, Type B
Environmental Limit
Conditions
(2)
Operating conditions: 5
o
C to 40
o
C
(41
o
F to 105
o
F ) at altitudes up to 10,000 ft.
(523mmHg) above sea level.
Storage: -20
o
C – 60
o
C (-4
°F – 140°F)
Relative Humidity: up to 95% RH
(non – condensing)
(1) Based on 21
°C (70°F) at a nominal operating pressure range of
0-7 kPa (0-1 psig back pressure). Maximum output pressure: 55 kPa
(7.98 psig)
(2) Operating outside of these operational specifications can limit the
concentrator's ability to meet Oxygen Concentration specification at
higher liter flow rates.