3
General description
3.1 Power supply
Depending on the circuit type, the XB transmitter requires the following power supply:
a) XB2 (2-wire, loop powered): 10…28 VDC - depending on the load connected to the output(s). The
minimum supply voltage can be determined as follows:
V min = 10 V + (0.02 x Load*) *Load resistance in ohms.
For the maximum load of 500
, the minimum supply voltage is 10 + (0.02 x 500) = 20 VDC. With both
output circuits closed, the maximum current consumption is 40 mA.
b) XB3 (3-wire with analog outputs): 15 to 40 VDC (see note below) or 12 to 28 VAC. With both output
circuits closed, the maximum current consumption is 50 mA.
Note: depending on the type of output signal, the XB3 will operate with the following minimum voltage
0…1 V outputs:
5 VDC or 5 VAC
0…5 V outputs:
10 VDC or 8 VAC
0…10 V outputs:
15 VDC or 12 VAC
0…20 mA or 4 …20 mA outputs:
6 VDC or 5 VAC with 0
load
15 VDC or 12 VAC with 500
load
3.2 Measured parameters
The XB transmitter measures relative humidity with a ROTRONIC Hygromer
®
IN1 capacitive sensor and
temperature with a Pt100 RTD.
3.3 Calculated parameters
Using the ROTRONIC HW4 software, the XB transmitter can be configured by the user to calculate either the
dew point or the frost point.
3.4 Analog output signals
XB2 and XB3
With the ROTRONIC HW4 software any of the two analog output signals can be made to correspond to one of
the following:
Relative
humidity
Temperature
Dew or frost point
Any output can also be disabled.
The scale of each analog output can be set within the numerical limits of -999.99 and 9999.99.
The D/A converters used to generate the analog output signals feature a 16-bit resolution.