7.3.2
Adjust Pt100
Reference Value Adjustment
1. Enter a defined reference value with a Pt100 simulator or record the current nominal temperature with
a reference instrument.
2. Enter the reference value determined in MENU > Pt100 > RefValue.
3. Select
"
in MENU > Pt100.
After successful adjustment, "OK" is shown in the display.
Reset Adjustment
The following procedure is used to reset adjustment to the factory setting.
1. Select
"
in MENU > Pt100
After successful resetting of the adjustment, "OK" is shown in the display.