36 Command Reference

Head Temperature Calibration Coefficients

Syntax Function
HeadTempCal.a0[=n][?]

Polynomial 0

th

order (Zero) coefficient, ~ 0

HeadTempCal.a1[=n][?]

Polynomial 1

st

order (Span) coefficient, ~ 0.01

HeadTempCal.a2[=n][?]

Polynomial 2

nd

order (Linearity) coefficient, ~0

HeadTempCal.a3[=n][?]

Polynomial 3

rd

order (Linearity2) coefficient, ~ 0

HeadTempCal.avg[=n][?]

Amount of averaging applied to measurement, ~40

HeadTempCal.avgBand[=i][?]

Only average when measurements remain within
this band, ~ 1°C

HeadTempCal.sensCode[=i][?]

=37 for 10K thermistor

HeadTempCal.coefType[=i][?]

=2 for polynomial type coefficients

SaveCal=973

Save these and other calibration parameters

Flow Sensor Calibration Coefficients

Syntax Function
FlowCal.a0[=n][?]

Polynomial 0

th

order (Zero) coefficient

FlowCal.a1[=n][?]

Polynomial 1

st

order (Span) coefficient

FlowCal.a2[=n][?]

Polynomial 2

nd

order (Linearity) coefficient

FlowCal.a3[=n][?]

Polynomial 3

rd

order (Linearity2) coefficient

FlowCal.avg[=n][?]

Amount of averaging applied to measurement, ~0

FlowCal.avgBand[=i][?]

Only average when measurements remain within
this band, ~ 0 l/m

FlowCal.sensCode[=i][?]

=21 for voltage output sensor

FlowCal.coefType[=i][?]

=2 for polynomial type coefficients

SaveCal=973

Save these and other calibration parameters

System Identification

Syntax Function
ID?

Returns a string containing instrument
identification, ie. DPM 973

IDN?

Returns only numeric portion of identifier, ie. 973

Saving Changes

Syntax Function
SaveCfg=973

Save configuration parameters

SaveCal=973

Save calibration parameters

SavePID=973

Save PID parameters