ORIS Settings
Syntax Function
UseOris.on[=i][?]
1 enables ORIS, 0 disables ORIS
UseOris.below[=n][?]
Temperature below which ORIS activates, °C
SaveCfg=973
Save these and other configuration parameters
Mirror Cleaning Preparation
Syntax Function
MinHeadRemovalTemp[=n][?]
Head and inner cooler are warmed above this temp
when Mirror Cleaning button has been pressed
prior to head removal, °C
SaveCfg=973
Save these and other configuration parameters
Automatic Mirror Check
Syntax Function
AMC.on[=i][?]
1 enables AMC, 0 disables AMC
AMC.cycleTime[=n][?]
AMC cycle time in minutes
AMC.heatTime[=n][?]
Time in mins to hold at high temperature during
Manual or Automatic Mirror Check
AMC.temp[=n][?]
Temperature to heat to during Manual or Automatic
Mirror Check, °C
AMC.dispHold[=i][?]
1 freezes DP display during Manual or Automatic
Mirror Check
SaveCfg=973
Save these and other configuration parameters
Pump Control Parameters (if available)
Syntax Function
Pump.on[=i][?]
1
Æpump On, 0Æpump Off
Pump.freq[=n][?]
Pump solenoid frequency, Hz
Pump.dutyCycle[=n][?]
Pump solenoid on/off duty cycle
SaveCfg=973
Save these and other configuration parameters
Analog Outputs
See Analog Outputs on page 20 for explanation of the parameters.
Syntax Function
Ana1.param[=i][?]
Parameter to track with analog output #1
Ana1.paramMin[=n][?]
Min Value
Ana1.paramMax[=n][?]
Max Value
Ana1.OutMin[=n][?]
Min Voltage
Ana1.OutMax[=n][?]
Max Voltage
Ana1.calGain[=n][?]
Gain calibration factor for D/A converter
Ana1.calZero[=n][?]
Zero calibration factor for D/A converter
Ana1.output=n
Use to set analog output when calibrating D/A
converter. Automatically sets Ana1.hold=1. You
must send Ana1.hold=0 when finished with
calibration.
Ana1.hold[=i][?]
1
ÆCalibration mode. Analog output is set with the
command Ana1.output=n.
0
ÆNormal mode. Analog output tracks the
parameter set with Ana1.param=i.
SaveCfg=973
Save these and other configuration parameters