Installation, Facility requirements, Environmental – ROTRONIC MBW 973 User Manual
Page 31: Power, Preparation for use, Benchtop use, Preparation for shipping or transportation, The installation se

Facility Requirements 27
Installation
Facility Requirements
Environmental
Your 973 is a precise laboratory quality instrument. While it does not require any special
environmental control, it works best when the temperature is stable and free of rapid transitions.
For operation, it is best when used under the following conditions:
Operating Temperature
15 to 35°C
Operating Humidity
5 to 95%RH non-condensing
Power
The 973 is equipped with a universal power supply capable of operation from 100 to 250 VAC at
50 to 60 Hz. If different, the specific power requirements will be identified on the back panel.
Preparation for Use
Prior to first measurement the instrument should be switched on for 30 minutes to warm up.
Benchtop Use
All 973 models are ready for benchtop use. Bench space of at least 24” x 24” (0.6 x 0.6 m) is
recommended.
Preparation for Shipping or Transportation
All shipping and transportation should be done in suitable padded containers. A heavy duty
container with at least 2 to 4 inches of clearance for foam padding is recommended.
Due to the design of the 973 systems, any shipping carton used should be constructed to
provide support only on the top and bottom of the side panels and on the front and rear frame of
the unit. Try not to have any load bearing directly on the face, nor on the top or bottom center of
the front or rear frame. Reusable shipping containers are available. Please contact the factory
for pricing and availability.
Prior to shipping, always cap the gas inlet and outlet fittings to prevent excess moisture and/or
contamination from entering the gas path, and to prevent damage to the fittings.
Ensure the optical head assembly and cover screw are firmly attached.