Dwyer 32A User Manual
Page 31
![background image](/manuals/371965/31/background.png)
Addr
(Option 992, Serial Communications) Control Address: Set from 1 to 3FF. This number
(hexadecimal, base 16) must match the address number used by the host computer.
bAUd
(Option 992, Serial Communications) Communication Baud Rate: Select 300, 1200, 2400,
4800, 9600, or 19200. This number must match the baud rate used by the host computer.
nAt
(Option 992, Serial Communications) No Activity Timer: Set from OFF or 1 to 99 minutes.
1 - 99
Maximum time between host computer accesses. If timer counts to 0, CHEC/LorE
will be displayed.
OFF
No Activity Timer function is disabled.
Stor
(Option 992, Serial Communications) Store to EEPROM: Select YES or no.
(See additional information on page 18).
YES
Menu Items changes made through the Serial Communications are stored
directly to the EEPROM.
no
Menu items changes made through the Serial Communications are stored
directly in RAM.
When Stor is set to no, ArUP must be set to ON to make sure no memory is stored.
Page 31
NOTICE