Dwyer GFC User Manual
Page 27

23
INDICATION
LIKELY REASON
REMEDY
lack of reading
power supply off
check connection of power supply
or output
fuse blown
disconnect transducer from
power supply; remove the
shorting condition or check
polarities; fuse resets
automatically
filter screen
flush clean or disassemble to
obstructed at inlet
remove impediments or replace
occluded sensor tube
flush clean or disassemble to
remove impediments or return to
factory for replacement
pc board defect
return to factory for replacement
valve adjustment wrong
re-adjust valve (section 7.3)
flow reading
inadequate gas pressure
apply appropriate gas pressure
does not
coincide with
filter screen obstructed
flush clean or disassemble to
the setpoint
at inlet
remove impediments or replace
ground loop
signal and power supply
commons are different
no response
inadequate gas pressure
apply appropriate gas pressure
to setpoint
cable or connector malfunction
check cables and all connections
or replace
setpoint is too low
re adjust setpoint or disable 2%
(<2% of full scale)
cutoff feature (section 2.2)
valve adjustment wrong
re-adjust valve (section 7.3)
unstable or
gas leak
locate and correct
no zero reading
pc board defective
return to factory for replacement
full scale output defective sensor
return to factory for replacement
at "no flow"
condition or
with valve
gas leak
locate and repair
closed
8.2 Troubleshooting Guide