4. SPECIFICATIONS

FLOW MEDIUM: Please note that GFC Mass Flow Controllers are designed to work with
clean gases only. Never try to meter or control flow rates of liquids with any GFC's.

CALIBRATIONS: Performed at standard conditions [14.7 psia (1.01 bars) and 70

0

F

(21.1

0

C)] unless otherwise requested or stated.

ENVIRONMENTAL (per IEC 664): Installation Level II; Pollution Degree II.

ACCURACY: ±1.5% of full scale, including linearity for gas temperatures ranging from
59

0

F to 77

0

F (15

0

C to 25

0

C) and pressures of 5 to 60 psia (0.35 to 4.1 bars).

REPEATABILITY: ±0.5% of full scale.

TEMPERATURE COEFFICIENT: 0.15% of full scale/

0

C.

PRESSURE COEFFICIENT: 0.01% of full scale/psi (0.07 bar).

RESPONSE TIME: GFC-110/111: 300ms time constant; approximately 1 second to

within ±2% of set flow rate for 25% to 100% of full scale flow.

GFC-113/114: 600ms time constant; approximately 2 seconds to
within ±2% of set flow rate for 25% to 100% of full scale flow.

GFC-1143/1144/1145:1800ms time constant; approximately 5
seconds to within ± 2% of set flow rate for 25% to 100% of full
scale flow.

GAS PRESSURE: 500 psig (34.5 bars) max; optimum pressure is 20 psig (1.4 bars).

MAX DIFFERENTIAL PRESSURE: 50 psid for GFC17/37/57/67/77, 40psid for GFC47.

GAS AND AMBIENT TEMPERATURE: 32

0

F to 122

0

F (0

0

C to 50

0

C).

RELATIVE GAS HUMIDITY: Up to 70%.

LEAK INTEGRITY: 1 x 10

-7

sccs He max to the outside environment.

ATTITUDE SENSITIVITY: No greater than ±15 degree rotation from horizontal to vertical;
standard calibration is in horizontal position.

OUTPUT SIGNALS: Linear 0 to 5 VDC (1000

Ω minimum load impedance) and 4 to 20

mA (50 to 500

Ω loop resistance); 20 mV peak to peak max noise for

GFC-110/111/113/114 and 100 mV peak to peak max noise for GFC-1143/1144/1145.

COMMAND SIGNAL: Analog 0 to 5 VDC (100

Ω input impedance) or 4 to 20 mA (0 to

250

Ω loop resistance).

Contact your distributor or Dwyer for optional RS232 or IEEE488 interfaces.

TRANSDUCER INPUT POWER: +12 VDC, 1100 mA maximum;
GFC-110/111/113/114 have an OPTION of +24 VDC, 650 mA maximum -
IF SPECIFIED AT TIME OF ORDERING AND CONFIGURED ACCORDINGLY.