In the Flow Chart:
1.
Each separate screen is inside a box.
2.
If a screen “branches”, an arrow points right and each screen of the branch is in a box.
3.
When there is text, but it’s not in a box, that signifi es a parameter setting and a short explanation is given of the parameter.
Mark on the chart what parameters need to be set. Note parameter settings on the chart. Program your amplifi er accordingly.
Keep the chart as a reference for other shifts and personnel and to monitor meter performance.
FLOW CHART - MENU STRUCTURE
Menu Structure
Meter Setup
Pipe Dia.
Enter the flow tube diameter
Detect Factor
Enter the meter calibration factor
Excitation Freq
Select the coil excitation frequency
Calibration
Electric Zero
All calibration factors
Elec Full Scale
are determined at the
Hydr. Zero-Man
factory.
Hydr. Zero-Auto
Empty Pipe
Cal. empty pipe
Empty pipe detection setup.
Cal. full pipe
Enable/Disable
Chg. Password
Password new
Enter 0000 to disable password.
Measurements
Rate Unit
Select flow rate unit
Totalizer Unit
Select total volume unit
FullScaleFlow
Enter full scale flow for the analog output.
LowFlowCutOff
Enter the low flow cut off. Typically 0.2%
Flow Dir.
Select uni-directional or bi-directional flow.
Damping Factor
Select as needed.