Badger Meter M-Series M3000 User Manual
Page 23
23
4. Press + fi ve times to position the cursor at Empty Pipe.
5. Select or press E.
6. Screen Empty Pipe 15 opens.
Empty Pipe 15 is a typical select ON/OFF screen.
Empty Pipe is shipped OFF. To turn Empty Pipe, On:
1. Press
+ once to move the cursor down one place,
to Cal. empty pipe.
2. Press
E.
3. In
screen
EMPTY CALIB. 1E, press + twice.
4. Press
E to turn Cal=[ON].
5. Press
+ to move cursor to Exit WITH save.
6. Press
E to lock the selection into the program.
Continue pressing E to reverse through the screens to
your next programming selection or to return to the Main
Screen.
You have now used the +, - and E selections to access the
three types programming screens, List, Input Numbers
and Select ON or Off and maneuver through some meter
programming.
What to Program
There are a variety of screens available. Refer to the "Flow
Chart - Menu Structure" on page 35 for all programming
options.
PROGRAMMING REQUIRED PARAMETERS
All meters have required parameters that must be
programmed. They include a Password (if desired), the Main
Screen for Uni-or Bi-Directional Flow, Empty Pipe, Pulse
Output and Analog Output, among others.
This section presents keystroke details describing how to
program required parameters.
Input a Password
1. With Main Screen up, press E, to open MAIN MENU 00.
2. With MAIN MENU 00 up, press + once to move cursor to
Meter Setup. Press E.
3. On METER SETUP 10 screen, press + six times or -
once to go to Chg. Password.
4. At CHG. PASSWORD 16 screen, input a password
number.
5. Or this screen comes up. Input the correct password.
6. If the wrong number is input, INVALID PSWD comes up.
Press +, - or E once.
7. Main Screen appears. Press E again.