Errors & Warnings
What You See
Why It Happened
How to Fix It
MEASURE_TIMEOUT
The board may be damaged .
Contact Badger Meter Technical Support .
COMMON_MODE_VOLTAGE_
OVERLOAD
Common mode voltage is smaller than
–2 .0V or larger than +4 .1V .
Make sure the meter is properly grounded .
Empty detector .
Switch Empty Pipe detection ON .
Dirty electrodes .
Clean the electrodes .
EMPTY_PIPE_WARNING
Measured impedance between the Empty Pipe
electrode and the Ground exceeded the set value .
• Make sure the pipe is filled .
• Check Empty Pipe threshold . It should be
60000 Ω (corresponds with 20 µS/cm) .
Empty detector .
PULSE_OVERLOAD_WARNING
Overflow occurred on the flow output .
Lower the pulse number .
FLOW_OVERLOAD_WARNING
Flow exceeded the Full Scale of more than 100% .
Set the flow range properly .
LOW_POWER_WARNING
Battery voltage is smaller than 3 .0V .
Consider replacing the battery .
EEPROM_ERROR
Configuration file is missing .
Replace the board .
CONFIG_ERROR
Configuration file is corrupted .
Replace the board .
PREAMPLIFIER_OVERLOAD
Input voltage exceeded the limits .
Maximum polarization is ± 227 mV; maximum
power line noise is 10 .6 mV; maximum useful
signal is 10 .7 mV .
Empty detector .
Switch Empty Pipe detection ON .
Dirty electrodes .
Clean the electrodes .
COIL ERROR
Coil/sensor is not connected .
• Make sure wiring is properly connected to the
amplifier .
• If that does not clear the alarm, contact
Badger Meter Technical Support .
Coil shorted .
Issue with coil current .
Troubleshooting