Descriptions of Simple Script Commands
Write Address

The Write Address command allows you to write a value to a register address

of the instrument. You are required to enter Execution Time, Address Code

and Address Value. The Address Code is the actual address to be changed,

while the Address Value is the value for the register.
Please note: Written register values are not easily undone. You must rewrite

the original value to the register if you want to undo your changes.

Please use this function only if you are sure you need to do so and only

after recording the original register value. If you are unfamiliar with register

values and their corresponding functions, please contact Alicat or your local

distributor for information.

Change Set-point (Scale Units)

The Change Set-point (Scale Units) command lets you change the set-point

of the instrument as a function of the unit’s scale. Selecting this command will

cause an Input window to appear that requires input of Execution Time (in

seconds) and Scale Units.
Please note: The Change Set-point (Scale Units) function is only available

with controllers.

Change Set-point (% of Full Scale)

The Change Set-point (% of Full Scale) command allows you to change the

set-point of the instrument as a function of percentage (0 - 100%). Selecting

this command will cause an Input window to appear that requires input of

Execution Time (in seconds) and % of Full Scale. It is not necessary to use the

“%” symbol when populating the Parameter Value.
Please note: The Change Set-point (% of Full Scale) function is only available

with controllers.

Tare

The tare function is used to set the zero on the instrument. Tare only when

there is

no gas flow going to the unit. The tare function is discussed in detail in

the instruction manual that accompanied your instrument. Choosing the tare

function will cause an Input window to appear that requires input of Execution

Time (in seconds).

Reset Total

The Reset Total function is used to reset the instrument’s flow totalizer.

Choosing the Reset Total function will cause an Input window to appear that

will require input of Execution Time.
Please note: The Reset Total function is available only for instruments ordered

with the optional totalizer.