Valve Tuning — If you require specialized valve tuning, you may do this via

the “Control” section in the “Advanced” section.
Adjustment of PD&I values will change how the valve responds, valve stability

and valve speed.
It is strongly recommended that adjustments to PD&I settings not be made

unless you have some experience with proportional valve tuning.

Adjusting The PD&I Settings Without Prior Training May Result In

Permanent Damage To Your Control Valve.

Valve offset or preload is a setting to compensate for the mechanical pressure

effects that gas pressure may exert on the control valve. Gas pressure in the

system exerts a mechanical force on the control valve that, depending upon

the gas pressure, may slow its response when opening.
By increasing valve offset, a signal is sent to the valve that will increase its

force when opening.
Adjusting valve offset may help your Alicat controller respond more quickly

when moving away from a zero set-point.
Valve offset is a value between 0 and 100, 0 being little to no offset and 100

being maximum offset.
If your controller uses an Alicat medium sized valve, the offset effect is reversed.

0 being maximum offset and 100 being little to no offset.
Valve offset has no effect at a zero set-point.
Info Tab

The info tab provides specific information about the unit.
This information is helpful when requesting telephone support or authorization