Page 47: 9 installation and calibration

background image

.

9

Installation and calibration

When the operating voltage is applied to SG45 after connecting e.g. the load cells, the unit starts running.
An automatic check for connection of the Sense lines takes place. This check is done also after configuration changes.
For installation, press

ô

during 3 sec to change over to installation mode

InSt

.

Select

SEt

ô

and adjust

S.typ (cell type) to select calibrating method CAL.M.

9.1

Initial setting (

SEt)

S.tyP for selecting the sensor type, see page 39.

CAL.M

shunt calibration

When connecting load cells, set

CAL.M= 0 (without shunt calibration).

CAL.M=1 (with shunt calibration Ü)

During calibration, the calibrating resistor at the sense connection

¡ i

s included automatically during calibration

of the upper calibration point (

InH.1 / OuH.1 )

1

Only purposeful for melt pressure sensors with integrated calibrating resistor. The specifications of the sensor
manufacturer related to the operating temperature must be taken into account..

2

Simulates a bridge load

Execute the other entries accordingly.

DP

decimal point

Out.1 analog output scaling
D.Unt unit of display shown on the 2nd display line
O.tyP output signal type OUT3

donE Subsequently, calibration ( CAL) takes place.

Installation and calibration

47

Initial setting (

SEt)

SG 45

Kg

SEt

CAL

S.tyP

InL.1

Out.1

CAL.M

InH.1

D.Unt

InL.1

InH.1

OuH.1

OuH.1

OuL.1

dp

O.tyP

SCAL

End

donE

donE

donE

InSt

45.6

3s

SEt see 9.1

CAL

see 9.2 Page 43

SCAL

see 9.3 Page 49