11.4.4

Control signal of Pulse length Controllers

Relay

ON

tON

tOFF

Pulse Length T

OFF

Time [s]

The output relay of the pulse length controller is clock-timed. The switching period T remains constant.
Depending on the divergence from the limit value, the switch on time tON is increased or decreased in
accordance with the proportional range Xp.

The following applies:

tON + tOFF = T (Const.)

greater divergence

Î greater tON

Xp exceeded

Î tON = T (relay remains picked up)