6 setup mode, 1 enter setup mode, Nter – Thermo Fisher Scientific Alpha pH 2000D User Manual
Page 21: Etup mode, Scd 22” if you want to change parameter settings, Press the ent key

6 Setup
Mode
6.1
Enter Setup mode
In the Setup mode the transmitter can be configured to your individual requirements.
1. While in Measurement mode press the ENT key.
2. The display prompts you to enter the security code. Set the security code with
▲ or ▼ key
to:
MEAS
pH
O
C
− “SCD 22” if you want to change parameter settings
ENTER
ENT
− “SCD 000” if you want to view only parameter settings (view only mode)
3.
Press the ENT key
.
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
ENTER
ENT
SETUP
HOL D
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
Temperature settings
see section 6.3
Relay A (set point 1) settings
see section 6.4
Relay B (set point 2) settings
see section 6.4
Controller settings
see section 6.5
Current output 1 settings
see section 6.6
Current output 2 settings
see section 6.7
Wash contact settings
see section 6.8
Unit settings
see section 6.9
Calibration
see section 5
Note: Note: to exit Setup mode at any time press the
▲ and ▼ key simultaneously
(escape). The transmitter returns to the Measurement mode.
- 16 -