6.7

Current Output 2 sub-function

In this sub-function you set the current output range of the transmitter for temperature values.

ENT

ENT

ENT

ENT

ENT

ENT

ENT

o

c

o

c

2

2

3

3

4

4

o

c

o

c

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

1

cur.2

SETUP

HOLD

1. Select the “CUR.2” sub-function, then press the ENT key.

2. Selecting the output type: press the

▲ or ▼ key to select the desired output type: 0-20 mA or 4-20

mA. Press the ENT key to confirm your selection.

3. Setting the temperature value at which the transmitter output will be 4 mA (or 0 mA): press the

or

▼ key to set the temperature value (setting range: -10.0 to 110.0 °C or 14 to 230 °F for NTC300

thermistor) to be equivalent to 4 mA (or 0 mA). Press the ENT key to confirm your setting.

4. Setting the temperature value at which the transmitter output will be 20 mA: press the

▲ or ▼ key

to set the temperature value (setting range: 0.0 to 110.0 °C or 32 to 230 °F for NTC300 thermistor) to be
equivalent to 20 mA. Press the ENT key to confirm your setting.

Continue with Setup mode procedures, or return to Measurement mode by pressing the

▲ and ▼ keys

simultaneously (escape).