Setting temperature (Set °C°F) sub-function

SETUP

HOLD

SETUP

HOLD

O

C

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

O

O

C

F

SETUP

HOLD

O

C

ATC

SETUP

HOLD

SETUP

HOLD

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1

2

3

3

4

5

6

1. Select the “SET °C°F” sub-function, then press the ENT key.

2. Selecting temperature unit: press the

▲ or ▼ key to select the desired temperature unit “°C” or “°F”.

Press the ENTER key to confirm your selection.

3. Enable/disable ATC mode: press the

▲ or ▼ key to enable (ATC ON) or disable (ATC OFF)

automatic temperature compensation. Press the ENT key to confirm your selection.

ATC disabled (ATC OFF):

For manual temperature compensation you can set the process and calibration temperatures.
This allows calibration at a temperature other than the process temperature. Example: setting a
calibration temperature of 25°C lets you calibrate using standard solutions at 25°C, even if your
process temperature is different from 25°C.

4. Setting the process temperature: press the

▲ or ▼ key to adjust the process temperature (upper

value, setting range: –10.0 to 125.0°C / 14.0 to 257.0 °F). Press the ENT key to confirm process
temperature.

5. Setting the calibration temperature: press the

▲ or ▼ key to adjust the calibration temperature

(upper value, setting range: – 10.0 to 125.0°C / 14.0 to 257.0 °F). Press the ENT key to confirm the
calibration temperature.

ATC enabled (ATC ON):

6. Setting the temperature offset: The display shows the current temperature offset (upper value) and the

temperature currently measured (lower value). Compare the current temperature displayed on the
transmitter to a thermometer known to be accurate. Note the correct temperature value. Press the

▲ or

▼ key to adjust the lower value. The upper number indicates the offset value. You can offset the
temperature by up to ±10°C/±18°F. Press the ENT key to confirm your setting.

Continue with additional Setup procedures, or return to the Measurement mode by pressing the

▲ and

▼ key (escape) simultaneously.