Thermo Fisher Scientific Alpha pH 2000D User Manual

Page 27

background image

2. Selecting the controller type: press the

▲ or ▼ key to select the suitable controller type:

OFF = controller off

Use control Off to operate controller as a monitor only or to prevent relays from switching.

L.CT = limit value control (on/off control).

Use limit control with pumps or valves for fast response

P/PI = proportional/integral control

Use proportional control to operate your pumps smoothly or for precise control of proportioning
valves. Use PI controller to eliminate steady state error.

Note: please refer to Appendix 4 for detailed information on controller settings.

Press the ENT key to confirm your selection.

If the controller is set to limit value control (L.CT):

3. Selecting the relay status under Non-Alarm condition: press the

▲ or ▼ key to choose the desired

relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.

If the controller is set to proportional control (P/PI):

4. Selecting the proportional controller type: press the

▲ or ▼ key to select the suitable controller type

(“PLC” = pulse length control, “PFC” = pulse frequency control). Press the ENT key to confirm your
selection.

If the proportional controller type is set to pulse length control (PLC):

5. Selecting the relay status under Non-Alarm condition: press the

▲ or ▼ key to choose the desired

relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.

6. Setting the proportional range: Press the

▲ or ▼ key to set the proportional range (setting range: 10

to 500%). Press the ENT key to confirm your setting.

7. Setting the pulse length: Press the

▲ or ▼ key to set the pulse length (setting range: 0.5 to 20

seconds). Press the ENT key to confirm your setting.

8.

Setting the integral action time: Press the

▲ or ▼ key to set the integral action time (setting range:

0.0 to 999.9 minutes). Press the ENT key to confirm your setting

.

If the proportional controller type is set to pulse frequency control (PFC):

9. Selecting the relay status under Non-Alarm condition: press the

▲ or ▼ key to choose the desired

relay status (de-energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.

10. Setting the proportional range: press the

▲ or ▼ key to set the proportional range (setting range: 10

to 500%). Press the ENT key to confirm your setting.

11. Setting the pulse frequency: press the

▲ or ▼ key to set the pulse frequency (setting range: 60 to

120 pulses per minute). Press the ENT key to confirm your setting.

12. Setting the integral action time: press the

▲ or ▼ key to set the integral action time (setting range:

0.0 to 999.9 minutes). Press the ENT key to confirm your setting.

Continue with Setup mode procedures, or return to Measurement mode by pressing the

▲ and ▼ keys

simultaneously (escape).