Current Output 1 sub-function

In this sub-function you set the current output range of the transmitter for pH or ORP values

.

2

2

3

3

4

4

5

5

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

pH

pH

pH

pH

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

1

cur.1

SETUP

HOLD

ENT

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1. Select the “CUR.1” sub-function, and then press the ENT key.

2. Selecting the output type: press the

▲ or ▼ key to select the desired output type: 0-20 mA or 4-20

mA. Press the ENT key to confirm your selection.

3. Setting the pH or ORP value at which the transmitter output will be 4 mA (or 0 mA): press the

or

▼ key to set the pH or ORP value to be equivalent to 4 mA or 0 mA, respectively. Press the ENT key

to confirm your setting.

4. Setting the pH or ORP value at which the transmitter output will be 20 mA: press the

▲ or ▼ key

to set the pH or ORP value to be equivalent to 20 mA. Press the ENT key to confirm your setting.

5. Selecting out of range current: Press the

▲ or ▼ key to switch “CU22” on or off. If “CU22” is on, the

output signal will jump to 22 mA when the measured pH and ORP value is below or above the values set
in points 3 & 4 above. Press the ENT key to confirm your setting.

Note: the “CU22” setting (on/off) of current output 1 applies to current output 2 as well.

Continue with Setup mode procedures, or return to Measurement mode by pressing the

▲ and ▼ keys

simultaneously (escape).