Wash relay (WASH) sub-function

In this sub-function you set the parameters for the wash relay.

1

2

2

3

3

4

5

ENT

ENT

ENT

ENT

ENT

ENT

ENT

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

SETUP

HOLD

off

1. Select the “WASH” sub-function, and then press the ENT key.

2. Enabling/disabling wash function: press the

▲ or ▼ key to enable (WASH ON) or disable (WASH

OFF) wash function. Press the ENT key to confirm your selection.

3. Selecting the relay status condition: press the

▲ or ▼ key to choose desired relay status (de-

energized = “DEEN” or energized = “EN”). Press the ENT key to confirm your selection.

4. Setting the wash interval in hours: Press the

▲ or ▼ key to set the desired wash interval (setting

range: 0.1 to 200.0 hours). Press the ENT key to confirm your setting.

5. Setting the wash duration in seconds: Press the

▲ or ▼ key to set the desired wash duration

(setting range: 1 to 2000 seconds). Press the ENT key to confirm your setting.

Note: During wash cycle the transmitter is set to HOLD. For safety reasons, the HOLD
function is activated 5 seconds prior and deactivated 10 seconds after the wash cycle.

Continue with Setup mode procedures, or return to Measurement mode by pressing the

▲ and ▼ keys

simultaneously (escape).











- 25 -