Techne TC-PLUS User Manual

Page 37

background image

35

temperature.

• Touch More Settings for further program defaults or

Back to return to the Settings menu.

Final Extension

Use this to add a final extension step to the program to

ensure complete amplification of the products.

• Touch On/Off to toggle between settings.

• Touch the time button to set the required hold time.

• Touch the temperature button to set the required

temperature.

Final Hold

This will hold the block at the set temperature at the

end of a program until the user manually stops the

unit.

• Touch On/Off to toggle between settings.

• Touch the temperature button to set the required

temperature.

• Touch More Settings to return to the top of the

Program Defaults menu or Back to return to the

Settings menu.

SERVICE SETTINGS

The service settings are for the use of authorised service engineers only. A service engineer’s

password is required to access these settings.

GRADIENT CALCULATOR

The Gradient Calculator can be used to calculate the temperature for each column of the block for a

user-specified temperature and gradient range.

• Touch the target temperature button and enter the

required temperature.


• Touch OK.


• Touch the gradient button and enter the required

gradient range followed by OK.

• Touch OK.

The calculated temperatures for each of the columns

of the block will now be displayed.

Note: With the 384-well block, the columns are paired

as it is not possible to individually control 24 columns.

Therefore only 12 column temperatures are available

across this block.