Techne TC-PLUS User Manual
Page 37
35
temperature.
• Touch More Settings for further program defaults or
Back to return to the Settings menu.
Final Extension
Use this to add a final extension step to the program to
ensure complete amplification of the products.
• Touch On/Off to toggle between settings.
• Touch the time button to set the required hold time.
• Touch the temperature button to set the required
temperature.
Final Hold
This will hold the block at the set temperature at the
end of a program until the user manually stops the
unit.
• Touch On/Off to toggle between settings.
• Touch the temperature button to set the required
temperature.
• Touch More Settings to return to the top of the
Program Defaults menu or Back to return to the
Settings menu.
SERVICE SETTINGS
The service settings are for the use of authorised service engineers only. A service engineer’s
password is required to access these settings.
GRADIENT CALCULATOR
The Gradient Calculator can be used to calculate the temperature for each column of the block for a
user-specified temperature and gradient range.
• Touch the target temperature button and enter the
required temperature.
• Touch OK.
• Touch the gradient button and enter the required
gradient range followed by OK.
• Touch OK.
The calculated temperatures for each of the columns
of the block will now be displayed.
Note: With the 384-well block, the columns are paired
as it is not possible to individually control 24 columns.
Therefore only 12 column temperatures are available
across this block.