SHIMPO 921A User Manual
Page 15

TRICOR Systems Inc.
2.5
Table 2-1
J2 Connector Wiring
PIN
SIGNAL
1
INSTALL
2
CH1 SENSE +
3
CH1 SENSE -
4
5
6
7
8
CH1 SRC+
9
CHASSIS GROUND
10
11
12
13
14
SIGNAL GND
15
CH1 SRC -
Notes:
1. Pin 9, chassis ground is provided for shield termination. If user-
designed nests have wire lengths in excess of 8 inches, those
wires should be shielded. If the user nest is to be used in an
electrically noisy environment, the wiring should be metal en-
closed with the metal enclosure chassis grounded.
2. Pin 1, INSTALL signal provides a device installed interlock for
testing. The test cycle will not initiate unless the Install signal is
grounded (connected to Pin 14). The user-designed nest can
incorporate a switch which is actuated when a device is in-
stalled in the nest, thereby grounding the Install signal. The
user's nest design can defeat the interlock by jumpering Pin 1 to
Pin 14 if the interlock function is not required.
3. Pin 3, CH1 SENSE - applies only to the 4
5resistance options.
- DT-361 (36 pages)
- DT-365E (28 pages)
- DT-311A (4 pages)
- DT-311J (12 pages)
- DT-315AEB (4 pages)
- DT-315P (17 pages)
- DT-326 (35 pages)
- DT-725 (4 pages)
- DT-900 (33 pages)
- ST-1000 (2 pages)
- ST-5000 (4 pages)
- SG-300 (2 pages)
- DT-721 (4 pages)
- DT-329 (32 pages)
- DT-105A (2 pages)
- DT-107A (2 pages)
- MT-100 (4 pages)
- MT-200 (4 pages)
- PT-110 (2 pages)
- DT-105A-S12 (2 pages)
- DT-205L (4 pages)
- DT-205L-S12 (4 pages)
- DT-207L (4 pages)
- DT-207L-S12 (4 pages)
- DT-209X (68 pages)
- DT-5TXR (20 pages)
- DT-5RL-1 (2 pages)
- DT-5TP (8 pages)
- DT-601CG (54 pages)
- FGV-XY-Series (32 pages)
- FGE-X (8 pages)
- FGE-HX (8 pages)
- FG-3000 (8 pages)
- FG-7000-Series (8 pages)
- DFS-series (20 pages)
- DRI-series (20 pages)
- EM-series (2 pages)
- FGE-series (4 pages)
- FGV-series (4 pages)
- FGE-H-series (4 pages)
- FGV-H-series (4 pages)
- MF Series (2 pages)
- MFD_Series (2 pages)
- FGS-100E (20 pages)
- FGS-100-PXL (7 pages)