10. Specification
¡Measuring Ranges and Accuracy
* Continuous recording mode [RMS] (50/60Hz,Sine wave)
Crest Factor ≦ 2.5 : Sine wave accuracy +2%+5dgt
* Event/Max. value/ Capture recording mode [RMS] (50/60Hz,Sine wave)
* Current detection (Event/ Max. value/ Capture recording mode):
* Capture recording mode [instantaneous value]:
NOTE: Electromagnetic compatibility(EMC)
EN61000-4-2 Electrostatic discharge immunity(ESD)
Performance criteria: B
¡
Operating system
:
Successive Approximation
¡
Input
:
AC voltage(AC100mV/A)
¡
Rated max. working voltage : AC170mVrms, 250mV peak value
¡
Number of input channel
:
3
¡
Measuring method
:
True RMS
¡
RMS Measuring interval
:
Continuous recording mode:
approx. 1sec. to 60 min. depends on recording Interval.
(intermittent sampling)
Max. value, Event mode
:
approx. 100ms. Normally, sampling at 3.3ms intervals. (Current
detection: at the interval of about 1.6ms.)
Capture recording mode
:
approx. 100ms. Consistently , sampling at 1ms intervals.
(Current detection: at the interval of about 0.5ms.)
Range
100mA
1000mA
0∼100.0mA
0∼1000mA
Measuring
range
Accuracy
of the instrument
Accuracy when
combining with sensor
±1.0%rdg±5dgt
±2.0%rdg±10dgt
±2.0%rdg±6dgt
Range
100mA
1000mA
0∼100.0mA
0∼1000mA
±1.5%rdg±7dgt
±2.5%rdg±12dgt
±2.5%rdg±8dgt
Range
100mA
1000mA
0∼100.0mA
0∼1000mA
±3%rdg±2%fs
±4%rdg±2.5%fs
±4%rdg±2%fs
Measuring
range
Measuring
range
Accuracy
of the instrument
Accuracy
of the instrument
Accuracy when
combining with sensor
Accuracy when
combining with sensor