KYORITSU 5000 Instruction Manual User Manual
Page 13

― 11 ―
Step2: Confirmation and change of set value
Confirm the mark indicating the
selected recording mode.
Refer to "5. Recording modes and
conditions
" in this manual to change
the recording mode or condition
(Recording interval/ Preset current
value).
Recording
mode
Details
Refer to:
Intermittent measured value is recorded
continuously at the preset interval.
(15-kind: 1 sec. to 60 min.)
P.16
When the preset current value is exceeded ,
(hereinafter, this event is referred as
current detection) three previous RMS
value, the RMS value when detecting it,
and four subsequent RMS values, 8 data in
total (0.8 sec.) are recorded.
P.19
The max. RMS value is recorded at every
10 sec. prior to and subsequent to the
current detection. It ends when a value
drops to 50% or less of the preset current
value; or when 10 min. has been elapsed.
P.21
Ten to twelve waveforms (for 200mS) are
recorded prior to and subsequent to the
current detection.
P.23
Continuous
recording
Event recording
Max. value
Recording
C a p t u r e
recording