KYORITSU 5000 Instruction Manual User Manual
Page 35

― 33 ―
7. "Timer": State of Timer function is displayed.
Can be set within "00:00" to "23:59".
¡
Press the
button and proceed to
the next step (Scale) when not changing
the setting.
¡
To change the setting, press the
button.
Then the indication on the LCD flashes.
Press the
or
button to set
the value to the desired one. Then press
the
button to confirm it.
Note
Press the "
" button at least 1 sec.
after setting. Then the instrument goes into
stand-by mode for recording. Recording
starts at the preset time.
8. "Scale": Scale value at Channel 1 is
displayed.
Can be set within "0.1" to "10.0". (1.0: OFF)
¡
Press the
button and proceed to
the next step (Scale value at Channel 2)
when not changing the setting.
¡
To change the setting, press the
button.
Then the indication on the LCD flashes.
Press the
or
button to set
the value to the desired one. Then press
the
button to confirm it.
Note
When measuring one-tenth signal via
Multi-tran or something, if the scale value is
set to 10.0; measured value can be directly
read from the LCD as it is displayed :
Measured value x 10.0 = Indicated value. (It
will not be reflected on the recorded data.)
* e.g.: Set "10.0" at CH 1. Then "150.0mA" is
displayed as a measured value when
measuring 15.0mA at CH 1.
፧
፧
፧
፧
፧