8-2-5 Setting for the interval between Auxiliary Earth Spikes at Earth

Resistivity (ρ) Measurements

Making setting of the intervals between auxiliary earth spikes is necessary
to measure earth resistivity (ρ).
Select “Lh” with the Cursor Key on the CONFIG_SETTING Screen, and
press the “

” Key to display the Length Setting Screen (Fig.16).

Select any digit to be changed with the Cursor Key, and press the “

Key. Then the selected digit is highlighted and ready to be changed. (Fig.17)

Press the Right Cursor Key

to increase numbers and the Left Cursor

Key

to reduce numbers. Keep the Cursor Key pressed down to change

numbers quickly. Press the “

” Key to confirm a number.

Repeat this procedure to change the other digits. Press the “

” Key

when settings are done. Then the CONFIG_SETTING Screen (Fig.18) with a
new interval will be displayed.

Note) Intervals can be set within a range of 1.0 to 30.0m. If a longer interval

out of this range is entered at the Setting Screen, it automatically
changed to “30.0m” when pressing the “

” Key.

Note) Intervals up to 20m can be selected with the supplied Test Leads

MODEL 7229.

8-2-6 Date and Time Setting
This instrument has a clock function and can save the measured data with
time and date information. The clock will not be reset once it has been set
even after powering off the instrument. A manual adjustment is required to
keep the clock time always right.

Fig. 16

Fig. 17

Fig. 18