● Dimension 167(L) x 185(W) x 89(D)mm

● Weight

approx 900g (including batteries)

● Power source DC12V : sizeAA manganese dry battery (R6P) x 8 pcs

* In a use of this instrument under low temperature below 0ºC, a use of

alkaline batteries with low temperature spec is recommended.

● Maximum Operating Error

  Operating error (B) is an error obtained within the rated operating

conditions, and calculated with the intrinsic error (A), which is an error of

the instrument used, and the error (En) due to variations.

          

)

15

.

1

(

2

5

2

4

2

3

2

2

E

E

E

E

A

B

+

+

+

+

±

=

A : Intrinsic error

E

2

: Variation due to changing the supply voltage

E

3

: Variation due to changing the temperature

E

4

: Variation due to series interference voltage

E

5

: Variation due to resistance of the probes and auxiliary earth electrode

resistance

● Range to keep the maximum operating error

  Measurement range within which the maximum operating error (±30%)

applies.

2Ω Range

: 0.5Ω ~ 2.099Ω

20Ω Range

: 2Ω ~ 20.99Ω

200Ω Range : 20Ω ~ 209.9Ω

2000Ω Range : 200Ω ~ 2099Ω

20kΩ Range : 2kΩ ~ 20.99kΩ

200kΩ Range : 20kΩ ~ 209.9kΩ

  Variation of Supply Voltage : until the Battery Warning mark appears

  Temperature Variation : -10℃ ~ 50℃

   Series Interference Voltage : 16・2/3Hz, 50Hz, 60Hz, 400Hz and

DC3V

  * No need to consider DC Series interference voltages at 2Ω & 20Ω

Ranges

   Auxiliary earth electrode resistance : within following range or 50kΩ

or less whichever smaller (At ρ measurements, "Re" in this formula is

replaced with "Rg".)

Rh,Rs limit

Accuracy

Re<0.40Ω

1kΩ

±5%rdg±1%fs

0.4Ω≦Re<1.00Ω

2kΩ

1.00Ω≦Re<2.00Ω

3.5kΩ

2.00Ω≦Re

=Rex100+5kΩ (Rh,Rs<50kΩ)