KROHNE BM 102 EN User Manual
Page 25
BM
102
25
Event
Fault
Action
Electrical Connections and Communication Output
No power supply
Check the power supply.
Connection of the device is
incorrect.
Check the connection between the
device and the power supply.
Current Output value
< 4 mA.
The calibration of the current
output is incorrect.
Execute the calibration if you have
authorized access or contact
KROHNE Service Centre**.
An error has occurred.
This happens in case the range 4-
20 mA / error 22 mA is selected.
Check the status of the device by
selecting the marker window (F8)
or enter the Watch status (4.0)
menu of the HART
®
communicator.
Reads 22 mA.
The device is in its start-up
phase.
Wait 50 seconds. If the current
value drops to a value between 4
and 20 mA, and goes immediately
back to 22 mA, contact your
KROHNE Service Centre**.
The value at the current
output does not correspond
to the value at the display
(PC STAR 2 or HART
®
communicator).
The current output settings
are incorrect.
Check the current loop and the
connections. Configure output as
described in Sect. 3.3.3 (user sub-
menu 1.3) of the BM 102
Handbook* - also try adjusting the
threshold using F11-Dynamic
Configuration function (PCSTAR 2)
or menu 2.1.5.1.0 – Threshold
(HHC).
The communication
parameters of the computer
are set incorrectly.
Check computer setting
(address/device number).
Bad connection to the
interface.
Check connection.
Current output value is < 4
mA.
Contact your KROHNE Service
Centre**.
Data communication via the
digital interface is not
working.
The BM 102 is in its start up
phase, wait 50 seconds and
try again.
Current output value is = 22
mA
If problem persists then contact
your KROHNE Service Centre**.