2 Electrical

connections

2.1

Electrical installation instructions

2.1.1 Wiring general notes

Connection to power is effected at the plug connector in the signal converter. Two wiring
compartments are available: DIN connector or M16 terminal box. Observe applicable rules and
regulations for cable connection: VDE 165 or equivalent national regulations.
Always disconnect from power supply before opening wiring compartment. This is not mandatory for
Exi applications.

Hazardous-duty systems


Only certified intrinsically safe equipment may be connected to
the BM 102 in hazardous areas. Please refer to the BM 102
MICROFLEX KEMA 00ATEX1101 X Supplementary Installation
and Operating Instructions for further information. This document
is available from your local KROHNE Sales office or on the
“Download Centre” on KROHNE’s website
http://www.krohne.com/.


2.1.2 Wiring

connections: DIN connector


Terminals:

3 poles and 1 x ground. Wire cross-section: max. 1.5 mm² (AWG 16)

Cable entry:

1 x PG11, cable diameter: 8 … 10 mm or 0.31” … 0.39”, IP 65 or NEMA4

Shielding:

do not connect to the DIN connector.

Signal cable:

no shielding required

1

Detach screw P and remove plug from signal
converter.

2

Separate part N from part R by inserting the flat
tip of a screwdriver into gap F.

3

Connect current loop to terminals 1 and 2 (any
polarity). Use ferrules to protect cable ends.
Terminals 3 and E are not connected.

4

Fit parts N and R together again.

5

Fit gasket, re-insert connector plug R on the
signal converter, and replace screw P.

Caution

Grounding terminal E is not connected to
the housing of the signal converter or to
the device flange. To avoid ground current
loops, the cable shielding may not be
connected at both ends.

Hazardous-duty systems

When used in hazardous areas, only one
intrinsically safe power supply may be
connected to terminals 1 and 2. Ground
terminal E and terminal 3 are not
connected.