MagCheck
5
1.3
MagCheck as manual signal calibrator
MagCheck replaces KROHNE Simulators GS8 / GS8A. With its integrated microprocessor,
calibrated mA-meter and frequency meter MagCheck allows for verifying the accuracy of converters
without the need for any further instruments like calculators, mA- and frequency meters.
MagCheck generates precise adjustable flow signals. The flow signal can be set in steps of 0.1% of
full scale. These very precise flow signals will be fed into the signal inputs of the converter. The
output signals (mA, frequency) are indicated on the MagCheck display. In this manual flow signal
calibrator mode no data will be stored in MagCheck. The measured/indicated data must be noted
and evaluated manually (see Appendix 1).
With non-IMoCom converters: In the flow signal calibrator mode with these previous converters data
as meter size (DN), GK, full scale setting must be entered into MagCheck manually.
MagCheck may be used
as flow signal calibrator for
following KROHNE
electromagnetic flowmeter
signal converters:
• T900
• SC100A, SC 100AS Non-Ex versions only!
• SC80A, SC 80AS Non-Ex versions only!
• AQUAFLUX
070
• IFC 080 Non-Ex versions only!
• IFC 200E, IFC 200. Non-Ex versions only!
• and all IMoCom converters as listed under 1.2
With IMoCom converters MagCheck loads data as meter size (DN), GK, full scale setting, current
and pulse output settings automatically down from IMoCom converter through the IMoCom
interface.