KROHNE MAGCHECK EN User Manual

Page 13

background image




MagCheck

13



3.2

Automatic verification with KROHNE IMoCom converters


3.2.1 How

to

start

verification


Menu Explanation

Display Details

1.0 Operation

Next screen

1.1 Test MID Starts automatic verification of flowmeter with

KROHNE IMoCom. All data will be verified and
stored in MagCheck.

inputs

xx/70: Number of stored verification files (e.g.
11/70), not transferred to the PC before. A total of
70 files can be stored. In this example 59
verifications can be made before memory is full.

date

Enter actual date (DD/MM/YY). This will allow PC
program to correctly assort verifications in trend
analysis. Attention: No plausibility check! Check
carefully!

Next screen

tag number

All signs on the MagCheck keyboard may be used
for entering the tag number. Method is the same
as typing an SMS on a cell phone.
Entered Tag Number will be used as identifier at
download into PC. You may use up to 16
characters.

Next screen measurement Test conditions must be defined

pipe full

filled with liquid
Electrode resistance will be tested via liquid.
With empty pipe the measured electrode
resistance will not comply with full pipe specs.
Result: Error statement in verification certificate

pipe empty

empty, clean, dry pipe
Electrodes will be tested for insulation. With a full,
wet or contaminated pipe, the measured electrode
resistance will not comply with empty pipe
specifications. Result: Error Message in
verification certificate

no pipe

Converter verification only
No verification of flow tube.

(Select acc.
to test /
process
conditions
using

↑↓.

This will
allow
MagCheck to
select the
correct
verification
program)

Process
liquid

Enter type, other data

Next screen

name user

Enter name of MagCheck operator

start?

Select "back" or "yes"

back

Will allow you to step back and check and correct
all data as entered above.

1.1
Test MID

(Select by
↑↓)

yes

Selecting yes and pressing

↵ will start automatic

verification run.