MagCheck



3

Description of MagCheck menu functions

3.1 Menu

structure


The menu of MagCheck has the following functional parts.

Menu Text

Description

1.0

Operation

Operation menu, main menu

1.1

Test MID

Automatic verification menu for electromagnetic flowmeters
with IMoCom- converters

..1.2 – 1.8

Menu for manual calibration test of converters

1.2

Parameter

Calibration test menu, entering meter data

→ Diameter

DN 2.5 – 3000 mm/0.1 – 120 inch

→ Full scale

Full scale value for flow Q=100%
(available display values: m3/hr, Liter/hr, USGal/min, l/s)

→ Primary constant See EMF tag plate

→ FCS nominal

Nominal field current value

1.3

Zero set

Zero setting

1.4

Simulation

Generation of flow signals, adjustable in 0.1%-steps of full
scale value

1.5

Curr. output

Display of true value (IMoCom converters only) and actual
value

1.6

Pulse output

Display of true value (IMoCom converters only) and actual
value

1.7

Field current

Indication of field current, displaying

• Positive

current

• Negative

current

• Sum of positive and negative current

• Field

frequency

1.8

Hardw. Info

Display of:

• Software ID of the respective modules

• Actual status information

(IMoCom converters only)

1.9

Reset

Reset menu (On IMoCom converters only)

2.0

Test

Information menu

2.1

Device Info

Display of MagCheck Serial number, software version and
calibration date

3.0

Install

Installation menu (main menu)

3.1

Language

Language setting. English, French, German

3.2

Data transf.

Data transfer menu from MagCheck to PC