MagCheck
Menu
Explanation
Display
Details, recommended actions
1.2 Parameter Menu for entering meter data.
→
Diameter
DN 2.5 – 3000 mm / 0.1 – 120 inch Select
using
↑↓.
↵
→
Full scale
Select physical unit using
↑↓.
(m3/hr, Liter/hr, USGal/min, l/s)
Enter full scale value for Q100% acc. type
plate or actual setting. In case of data
outside of meter range max / min values
are shown.
↵
Primary
constant
Enter GK / GKL acc. type plate. Take care,
no plausibility check.
↵
Enter nominal field current [mA]:
IFC 010
125
IFC 020, 090, 110, 210:
250
SC80, IFC 080, 200
125
1.2
Parameter
MagCheck starts
with KROHNE
standard
converter default
settings
Actual
parameters
listed right acc.
to meter setting
must be entered
manually acc. to
flow meter type
plate.
FCS nominal
T900 000
↵
1.3 Zero set
→
1.3
Zero set
Allows
MagCheck zero
adjustment
Physical unit as
set under 1.2
Default
000.000
For standard tests use default.
For precision tests follow 3.4.2.
1.4 Simulation
→
1.4
Simulation
Verifies
calibration and
linearity of ADC,
current (1.5),
pulse output
(1.6) by a
precision flow
signal of Full
scale as under
1.2
True value
in % + phys.
units)
Enter desired flow value (%).in 0.1% steps
Read converter display, note % and
flowrate.
↵
1.5 Current
output
→
1.5
Current
output
Shows current
output value at
flow rate (%) as
entered under
1.4
Actual value
[mA]
Read, note mA values.
↵
1.6 Pulse
output
→
1.6
Pulse output
Shows pulse
output value at
flow rate (%) as
entered under
1.4
Actual value
[Hz]
Read, note indicated pulse [Hz] rate.
↵
1.7 Field
current
→
1.7
Field current
Shows field
current peak to
peak (FCS),
pos., neg.
Act. value [mA] Note act. values [mA].
↵
1.8
Hardw. info
1.9
Reset
• In this operation mode menus 1.8 and 1.9 have no function.