TECHNICAL DATA

7

35

OPTISENS AAS 2000

www.krohne.com

12/2008 • MA AAS 2000 R01 en

Installation conditions

AAS 2000 + MAA 2000 fibreglass telescopic rod for
submersible installations

Installation on the handrail with up to 4 m
length-adjustable, oscillating fibreglass
assembly

Handrail mounting for:

- Round handrails: d = 32...50 mm / 1.3…2"

- Square cross-sections: 28...42 mm / 1.1…1.7"

AAS 2000 + MAA 2000 slide rail mounting for side wall
installations

Installation on side walls of channels and
basins using slide rails for simple sensor
removal.

Automatic spray cleaning

Flushing using clean water or compressed air;

Pressure: 2...6 bar / 29…87 psi

Solenoid valve: available in 220 V and 117 V
versions; up to max. 2 sensors can be operated
on a single valve

Flush hose: ¼" external diameter; PE; standard
length: 10 m / 32.8 ft

Process connection

Submersible version in open basins and
channels

Materials

Enclosure

SIS2343 (316 SS)

“Clark” electrode

Design: replacable cartridge

Cathode: gold

Anode: silver

Membrane: FEP Teflon (0.025 mm / 0.001")

Connection cable to converter

Insulation: Hytrel; 5-pin M 12 connector; fixed
cable, shielded, 10 m / 32.8 ft long

Flush hose

PE