beautypg.com

Operation – KROHNE OPTISENS AAS 2000 EN User Manual

Page 23

background image

OPERATION

5

23

OPTISENS AAS 2000

www.krohne.com

12/2008 • MA AAS 2000 R01 en

Menu "Calibrate"

Menu "Cleaning"

Menu "Scale / Alarm"

Submenu

Description

Airpres.mB

Airpres.mB

Airpres.mB

Airpres.mB

Specify within 10 mBar.

Calibrate

Calibrate

Calibrate

Calibrate

"No", "Zero" or "Air". Select what calibration to do.

Temp(info)

Temp(info)

Temp(info)

Temp(info)

Shows actual temperature in the electrode.

Test(info)

Test(info)

Test(info)

Test(info)

Shows actual dissolved oxygen level.

Setvalue

Setvalue

Setvalue

Setvalue

Theoretical reading in air at current temperature and with the air
pressure entered above. Shall be the same as Test

Test

Test

Test immediately after

the calibration.

Slope

Slope

Slope

Slope

Slope of the electrode at last air calibration.
This value is an indication of the wearing of the electrode and it is
normally over 80 for a well functioning electrode. However, an
electrode may also function well at lower values. At values below 50 it
is recommended to replace the electrode.

Calibrated

Calibrated

Calibrated

Calibrated

Date of last air calibration.

New electr.

New electr.

New electr.

New electr.

Date when electrode was last changed. Push ^ to edit the date, then
^ again to store it.

Submenu

Description

Press ^ to go to the cleaning program

Cleaner

Cleaner

Cleaner

Cleaner

"None", "Flush", or "Brush". Do not select "Brush" since this does
not exist for this sensor (only for master).

Interval min

Interval min

Interval min

Interval min

0...999 minutes, time between cleaning cycles (only for master)

Length sec

Length sec

Length sec

Length sec

0...999 seconds, duration of flushing cycle (only for master)

Freeze sec

Freeze sec

Freeze sec

Freeze sec

0...999 seconds, extra freeze time of output signal after a
flushing cycle

Relay

Relay

Relay

Relay

"-", "1", "2", "Along 1", or "Along 2". Select relay to operate solenoid
for flush cycle if this sensor is a master with its own relay, or relay
used by master if this sensor is a slave.
These are the same relays used for Alarm relay

Alarm relay

Alarm relay

Alarm relay below.

Next time

Next time

Next time

Next time

The next scheduled cleaning time. Pushing ^ on this line will set the
time to current time and start a cleaning cycle. This could be used to
test the "Flush" cycle (only for master).

Submenu

Description

Max

Max

Max

Max

0...99.9 mg/l, equal to 20 mA output signal.

Min

Min

Min

Min

0...99.9 mg/l, equal to 4 mA output signal.

Hi-Alarm

Hi-Alarm

Hi-Alarm

Hi-Alarm

0...99.9 mg/l, the value zero inactivates the alarm.

Low-Alarm

Low-Alarm

Low-Alarm

Low-Alarm

0...99.9 mg/l, the value zero inactivates the alarm.

Alarm Relay

Alarm Relay

Alarm Relay

Alarm Relay

"-" "1", "2", or "1 and 2". Check that the relay is not being used for
cleaning.

.book Page 23 Monday, December 15, 2008 12:43 PM