Operation, 4 scaling – KROHNE OPTISENS AAS 2000 EN User Manual
Page 27
OPERATION
5
27
OPTISENS AAS 2000
www.krohne.com
12/2008 • MA AAS 2000 R01 en
Air pressure conversion
It is important to enter the correct air pressure before doing an air calibration. Air pressure is
measured in mbar (which is exactly the same as the SI-unit hPa).Below is a formula and a table
to convert from inHG to mbar (and thus to hPa):
mbar = (inHg * 1000) ÷ 29.5
5.4 Scaling
The Scale / Alarm
Scale / Alarm
Scale / Alarm
Scale / Alarm menu (see the OPTISENS MAC 080 manual) allows the user to set the high and
low boundaries for a 4...20 mA output signal. In addition, this menu allows the user to set high
and low alarm values to switch a relay when solids have reached critical points.
inHg
inHg
inHg
inHg
mbar
mbar
mbar
mbar
29.2
990
29.3
993
29.4
997
29.5
1000
29.6
1003
29.7
1007
29.8
1010
29.9
1014
30.0
1017
30.1
1020
30.2
1024
30.3
1027
30.4
1030
Max
Max
Max
Max
sets the 20 mA point output
Min
Min
Min
Min
sets the 4 mA point output (may be negative for special applications)
Hi-Alarm
Hi-Alarm
Hi-Alarm
Hi-Alarm
sets the high alarm set point; the value zero inactivates the alarm
Low-Alarm
Low-Alarm
Low-Alarm
Low-Alarm
sets the low alarm set point; the value zero inactivates the alarm
.book Page 27 Monday, December 15, 2008 12:43 PM