INSTALLATION

3

13

OPTISENS AAS 2000

www.krohne.com

12/2008 • MA AAS 2000 R01 en

Installing a new electrode (see figure above)

Untighten the flushing head about one turn counterclockwise (if not already loose) to let the

air out when the electrode is pushed in place 1.

Take the electrode out of the plastic container.

Plug the electrode cable of the sensor into the connector on the electrode 2.

Install the new electrode into the flushing head 3.

Tighten the flushing head.

Mount the protection plate, if needed 4. See below for instructions on mounting the

protection plate.

Perform a a new air-calibration. An air calibration must be performed whenever the electrode

is changed.

i

Installation of the electrode is completed.

CAUTION!
Do not ever clamp onto the sensor housing or part of the cable when installing or removing the
flushing head.
Do not touch the membrane of the electrode when mounting.

Figure 3-3:

Installing a new electrode