Operation
5.1 Sensor display
By simultaneously pressing ↓ and ^ you alter between the converter main menu and the
sensor information display for the selected sensor.
The sensor has two information displays. The first one, in addition to the measured value, shows
the current temperature of the electrode, the setvalue for air and the "slope" of the electrode at
the last air calibration. The second information display shows the date of the last calibration and
the date the electrode was changed.
5.2 Menu
Use ↑ or ↓ to select the sensor in the main display. The menu for the selected sensor is accessed
by pressing ^ for five seconds. If the selected sensor is not active (the text No transmitter
No transmitter
No transmitter
No transmitter is
shown) a warning is displayed that asks you to make another choice in order to show the sensor
menu.
Menu "Settings"
Submenu
Description
Tag
Tag
Tag
Tag
Name of the sensor shown in the main display (10 characters).
I-Time
I-Time
I-Time
I-Time
Integration time or dampening – can be set up to 999 seconds.
Decimals
Decimals
Decimals
Decimals
1 or 2 for display and menu.
Analog
Analog
Analog
Analog
"None" , "Out1", "Out2", "Out3", "Out4", "Out1+2" or "Out3+4". Pick
which analog output(s) are to be used with sensor
Second
Second
Second
Second
"Temp" or "=Prim". If two analog outputs are chosen above, the first
will always give the primary value according to the sensors selected
scale. The second will either give the temperature scaled as stated
below, or the same signal as the first channel. The temperature is
additional information, not a precision measurement.
Temp 0-
Temp 0-
Temp 0-
Temp 0-
Temperature scaling, value that will give max output on second
channel.