Operation

5.1 Sensor display

By simultaneously pressing ↓ and ^ you alter between the converter main menu and the
sensor information display for the selected sensor.
The sensor has two information displays. The first one, in addition to the measured value, shows
the current temperature of the electrode, the setvalue for air and the "slope" of the electrode at
the last air calibration. The second information display shows the date of the last calibration and
the date the electrode was changed.

5.2 Menu

Use ↑ or ↓ to select the sensor in the main display. The menu for the selected sensor is accessed
by pressing ^ for five seconds. If the selected sensor is not active (the text No transmitter

No transmitter

No transmitter

No transmitter is

shown) a warning is displayed that asks you to make another choice in order to show the sensor
menu.

Menu "Settings"

Submenu

Description

Tag

Tag

Tag

Tag

Name of the sensor shown in the main display (10 characters).

I-Time

I-Time

I-Time

I-Time

Integration time or dampening – can be set up to 999 seconds.

Decimals

Decimals

Decimals

Decimals

1 or 2 for display and menu.

Analog

Analog

Analog

Analog

"None" , "Out1", "Out2", "Out3", "Out4", "Out1+2" or "Out3+4". Pick
which analog output(s) are to be used with sensor

Second

Second

Second

Second

"Temp" or "=Prim". If two analog outputs are chosen above, the first
will always give the primary value according to the sensors selected
scale. The second will either give the temperature scaled as stated
below, or the same signal as the first channel. The temperature is
additional information, not a precision measurement.

Temp 0-

Temp 0-

Temp 0-

Temp 0-

Temperature scaling, value that will give max output on second
channel.