6.5 Measuring accuracy

The measuring accuracy for partly filled pipes and completely filled pipes are different. In these
graphs it is assumed that the velocity at full scale value is at least 1 m/s (is also the standard
value for calibration, since it will result in the most accurate measurements).

Partly filled:

• v @ Full Scale ≥ 1 m/s / 3.3 ft/s: ≤ 1% of Full Scale

Fully filled:

• v ≥ 1 m/s / 3.3 ft/s: ≤ 1% of MV
• v < 1 m/s / 3.3 ft/s: ≤ 0.5% of MV + 5 mm/s / 0.2 inch/s (see following graph)

Fully filled pipes

Figure 6-3: Maximum measuring error of measured value