TECHNICAL DATA

38

TIDALFLUX 2300 F

www.krohne.com

06/2013 - 4002490202 MA TIDALFLUX 2300 F R02 en

6.3 Sizing

Flow and velocity at 100% level, gravity fed, metric values

INFORMATION!
These tables state the flowrate in a partially filled pipe, dependant on slope, filling level, inner
diameter and friction factors of the pipe. The values are calculated according the Manning-
Strickler equation. The results do not consider: deposits, profile distortion, backwater, friction of
air and the slightly smaller inner diameter of the TIDALFLUX.

DN

Steel / PVC

Smooth concrete

Concrete

Nominal

size

v [m/s]

Q [m3/h]

v [m/s]

Q [m3/h]

v [m/s]

Q [m3/h]

slope 0.5%

200

0.96

109

0.86

97

0.58

66

250

1.11

196

1.00

177

0.67

118

300

1.26

321

1.13

288

0.75

191

350

1.39

481

1.25

433

0.84

291

400

1.52

688

1.37

620

0.91

412

500

1.77

1251

1.59

1124

1.06

749

600

2.00

2036

1.8

1832

1.2

1221

700

2.21

3062

1.99

2757

1.33

1843

800

2.42

4379

2.18

3945

1.45

2624

900

2.62

6000

2.35

5382

1.57

3596

1000

2.81

7945

2.53

7153

1.68

4750

1200

3.17

12906

2.85

11603

1.9

7736

1400

3.51

19451

3.16

17511

2.11

11693

1600

3.84

27794

3.45

24971

2.3

16647

slope 1.0%

200

1.36

154

1.22

138

0.81

92

250

1.57

277

1.42

251

0.94

166

300

1.78

453

1.6

407

1.07

272

350

1.97

682

1.77

613

1.18

409

400

2.15

973

1.94

878

1.29

584

500

2.5

1767

2.25

1590

1.5

1060

600

2.82

2870

2.54

2585

1.69

1720

700

3.13

4336

2.82

3907

1.88

2605

800

3.42

6189

3.08

5573

2.05

3709

900

3.7

8474

3.33

7626

2.22

5084

1000

3.97

11225

3.57

10094

2.38

6729

1200

4.48

18240

4.03

16408

2.69

10952

1400

4.97

27542

4.47

24771

2.98

16514

1600

5.43

39302

4.89

35394

3.26

23596